Chapter 6 Troubleshooting
(14)
Attach a sterile syringe to the suction channel side of the injection tube and aspirate sterile
water with the syringe. While pressing the suction valve, flush sterile water into the instrument/
suction channel until the solid materials clogging the channel are removed.
(15)
Remove the weight of the injection tube from the clean basin filled with sterile water.
(16)
Aspirate air with the syringe. While pressing the suction valve, inject air into instrument/suction
channel until the sterile water inside the channel is discharged completely.
(17)
Detach the sterile syringe from the suction channel side of the injection tube.
(18)
Detach the tube for instrument/suction channel of the injection tube from the suction connector
on the LG connector.
(19)
Inspect the endoscope to check that it is free from abnormalities.
“Chapter 4 Preparation and Inspection”