<Portable and Mobile RF Communications Equipment Compliance Information and Guidance>
Table 7
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and users of
this product are advised to check that it is being used in such environments.
Immunity test
IEC 60601
Test level
Compliance level
Guidance
Conducted RF
IEC/EN 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF
communications equipment should
be used no closer to any part of this
product, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2
√
P
Radiated RF
IEC/EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
√
P
80 MHz to 800 MHz
d = 2.3
√
P
800 MHz to 2.5 GHz
Where “P” is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and “d”
is the recommended separation
distance in meters (m). The
electric field intensity from a fixed
RF transmitter determined by an
electromagnetic field study
a
should
be lower than the compliance level
in each frequency range
b
.
Interference may occur near devices
that display the following symbol.
Note
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which this product is used
exceeds the applicable RF compliance level above, this product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
this product.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
ED580XT_E2-30_897N200877B.indb 14
2018/09/20 15:50:26
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...