Chapter 6
Trouble
shooting
Problem
Cause
Remedy
Distorted image
High-frequency interference.
Stop power supply to the high-frequency
endotherapy device to restore image output.
The endoscope is working properly.
The endoscope is not connected
correctly to the light source.
Connect the endoscope properly to the light
source.
“4.6 Connecting Endoscope to Light
S o u r c e , P r o c e s s o r a n d R e l a t e d
Equipment”
6.1.2 Problem with Scope Switch
Problem
Cause
Remedy
The intended
function is not
executed even if
the scope switch
is pressed.
The intended function is not
assigned to the scope switch.
Assign the function to the scope switch as
per instructions provided in the Operation
Manual of the processor.
Operation Manual of the processor
6.1.3 Problem with Bending Section
Problem
Cause
Remedy
Bending section
cannot return to
neutral position.
T h e u p / d o w n a n d l e f t / r i g h t
angulation knobs are locked.
Turn the up/down and left/right angulation
locks in the direction of F (Free) until they
stop to unlock the up/down and left/right
angulation knobs.
ED580XT_E2-30_897N200877B.indb 3
2018/09/20 15:50:25
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...