897N200877B
5-5
Chapter 5
How to Use
(2)
Unlock the bending section by turning the up/down and left/right angulation locks in the
direction of F (Free) until they stop.
Note
The procedure described above is recommended. However, other techniques are also
available: you can insert endoscope by locking the bending section only in the left-right
direction and unlocking it in the up-down direction.
(3)
Raise the forceps elevator lever to lower the forceps elevator.
(4)
When necessary, wipe the objective lens and light guides with sterile gauze moistened with
alcohol.
Note
Use lint-free sterile gauze to prevent fibers from entering the air/water nozzle.
(5)
Apply clean lubricant (Xylocaine jelly or the like) to the insertion portion.
(6)
Turn on the light of the light source.
Note
If the light source is off, press the power button to turn it on.
(7)
Set the air supply pump’s operation of the light source to “HI”, “MID” or “LOW.”
(8)
Insert the distal end of endoscope into the oral cavity and then move it down to the pharynx
while under constant observation.
ED580XT_E2-30_897N200877B.indb 5
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...