897N200877B
5-9
Chapter 5
How to Use
5.2.5 Operating Air/Water Valve
Operate the air/water valve to supply air or water from the air/water nozzle in the distal end
during a procedure.
Note
Do not operate the air/water valve under the following conditions while the endoscope
is inserted in the body cavity. Doing so may cause the backflow of body fluid and
contaminate the sterile water in the water tank.
- The air supply pump of the light source is set to “OFF”.
- The connector of the water tank is not attached to the water supply connector of the
endoscope.
- The video connector of the endoscope is not connected to the scope connector
socket of the processor.
(2)
(1)
When the hole in the air/water valve is covered with one’s finger, air is supplied to the air/water
nozzle in the distal end.
(2)
When the air/water valve is pressed, water is supplied to the air/water nozzle in the distal end.
Note
• If patient material adheres to the distal objective lens or if the image is obscured,
clean the surface of the lens by operating the air/water valve and supplying water.
• If any debris such as mucus is left adhered, or if air is supplied without supplying
water, the debris may become hard to remove due to drying or fixation.
• After operating the air/water valve, if the endoscopic image is unclear due to the light
reflected off the water droplets remaining on the objective lens or distal cap, it may
be improved by performing suction.
ED580XT_E2-30_897N200877B.indb 9
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...