897N200877B
5-11
Chapter 5
How to Use
5.3 Injecting Fluids from Instrument Channel Inlet
WARNING
• When injecting fluids by attaching a syringe to the forceps valve, open the lid of the
forceps valve and insert the syringe straight into the forceps valve. Otherwise, the forceps
valve may be damaged or the syringe may be accidentally detached during fluid injection
and body fluids may leak or splash from the forceps valve, posing an infection control risk
to the patient or end-user.
• When the lid of the forceps valve needs to be opened during a procedure, place a piece of
gauze, etc. over it to prevent leakage. Otherwise, body fluids may leak or splash from the
forceps valve, posing an infection control risk to the patient or end-user.
Note
The lid of the forceps valve should normally be kept closed. When attaching the
syringe to supply water or fluid, remove this lid.
(2)
(3)
(4)
(1)
Fill a syringe with water or fluid.
(2)
Open the lid of the forceps valve.
(3)
Attach the syringe straight to the forceps valve and inject water or fluid.
(4)
Detach the syringe from the forceps valve and close the lid of the forceps valve.
ED580XT_E2-30_897N200877B.indb 11
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...