897N200877B
4-23
Chapter 4
Prepara
tion and Inspectio
n
4.6.2 Connecting to Processor
WARNING
• Firmly connect the video connector of the endoscope and the processor. If the video
connector is not connected properly, the endoscopic image may flicker or be lost, which
may cause patient injury, bleeding and/or perforation.
(1)
(2)
(1)
Align the index on the video connector and that on the 500 system connector socket of the
processor, and insert the video connector into the processor.
(2)
Rotate the lock handle clockwise while pressing the video connector further into the processor
to secure the connector.
(3)
Attach a dedicated connector cap to the connector socket that is not in use.
Light Source
Processor
Water Tank
4.6 Connecting Endoscope to Light Source, Processor and Related Equipment
Suction Unit
ED580XT_E2-30_897N200877B.indb 23
2018/09/20 15:50:23