897N200877B
5-3
Chapter 5
How to Use
5.1 Preparation
5.1.1 Preparing Related Equipment
Prepare related equipment to be used, including back-up or spare endoscope, scope
components (valves, distal caps, etc.) and accessories (ex. water tank, endotherapy devices,
etc.)
5.1.2 Pretreatment of Patient
Prepare the patient in the normal endoscopy regimen.
5.1.3 Preparing Mouthpiece
Prepare a mouthpiece per manufacturer’s instructions.
WARNING
• Use an appropriate mouthpiece as per manufacturer's instructions, including specific
reprocessing recommendations.
CAUTION
• Do not use a mouthpiece that is damaged, deformed, or reveals other irregularities. Doing
so may cause injury in the oral cavity and/or equipment failure.
Note
Reusable mouthpieces are considered consumable items. If any abnormality is found,
use a new mouthpiece per manufacturer's instructions.
(1)
Before oral insertion, have the patient hold the mouthpiece in his/her mouth.
Note
If you choose to have the patient hold the mouthpiece after insertion, attach the
mouthpiece to the insertion portion in advance. Have the patient hold it promptly after
insertion.
ED580XT_E2-30_897N200877B.indb 3
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...