Chapter 4 Preparation and Inspection
<Attachment>
(1)
Attach the forceps valve to the instrument channel inlet of the endoscope.
4.4.2 Suction Valve
Skirt
Spring
Hole
Suction Valve (SB-605)
WARNING
• Use a properly reprocessed suction valve. Use of an improperly reprocessed suction valve
could be a source of infection.
CAUTION
• When attaching the suction valve to the suction valve cylinder of the endoscope, align
the recessed and projecting portions and slowly insert the suction valve straight into the
suction valve cylinder of the endoscope. If the suction valve is attached forcibly, it may be
damaged.
<Preparation>
The suction valve is not reprocessed before shipping from FUJIFILM. Reprocess it according
to the instructions given in the Reprocessing Manual before using it in a procedure.
Reprocessing Manual
ED580XT_E2-30_897N200877B.indb 14
2018/09/20 15:50:22