5.6 Endoscope Withdrawal
(3)
(4) (5)
WARNING
• Unless the endoscope's bending section is in a neutral position (essentially "straight"), do
not withdraw the endoscope whenever the endoscopic image is compromised, unclear,
blurry, etc. due to any reason or condition including loss of image, power interruption,
water droplets or dirt/debris adhering to the objective lens, etc. Doing so may cause injury
to tissues in the body cavity, bleeding and/or perforation.
CAUTION
• Lower the forceps elevator before withdrawing the endoscope. Not doing so may cause
patient injury.
(1)
Prior to withdrawal, press the suction valve to apply suction to remove insufflated air (or CO
2
gas) from the body.
(2)
Raise the forceps elevator lever to lower the forceps elevator.
(3)
Prior to withdrawal, operate the up/down and left/right angulation locks in the direction of F (Free)
until they stop.
(4)
Prior to withdrawal, operate the up/down and left/right angulation knobs to straighten the
bending section to its neutral position.
(5)
Slowly withdraw the endoscope under constant visualization.
(6)
Turn off the light of the light source.
ED580XT_E2-30_897N200877B.indb 18
2018/09/20 15:50:25
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...