Chapter 5
How to Use
(4)
When the cannulation tube is advanced and contacts the forceps elevator and is unable to
advance any further, slowly raise the forceps elevator lever and insert the cannulation tube by
approximately 30mm.
Note
When the forceps elevator lever is slowly raised, the tip of the cannulation tube comes
into the field of view from the right lower portion of the screen.
(5)
Consistent with clinical training, direct the tip of the cannulation tube to the papilla opening and
insert the cannulation tube into the common bile duct or the pancreatic duct through the papilla
opening.
Note
To move the tip of the cannulation tube to the papilla opening, advance and withdraw
the cannulation tube by operating the anglulation knobs and the forceps elevator lever.
(6)
Inject a small amount of contrast agent with a syringe and check that the cannulation tube has
been inserted into the intended common bile duct or pancreatic duct.
(7)
Slowly inject the contrast agent.
(8)
Observe the conditions of the common bile duct or the pancreatic duct under fluoroscopy.
ED580XT_E2-30_897N200877B.indb 13
2018/09/20 15:50:25
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...