Chapter 4 Preparation and Inspection
4.3 Inspecting Endoscope
Prior to inspection, make sure that dirt does not adhere to the endoscope. If dirt adheres to the
endoscope, reprocess it according to the instructions given in the Reprocessing Manual.
4.3.1 Inspecting Control Portion
(1)
Visually and manually inspect the control portion for excessive scratches, deformation, cracks,
residue, loose parts, or other irregularities.
4.3.2 Inspecting Insertion Portion
CAUTION
• Do not forcibly twist or bend the insertion tube of the endoscope. It could damage the
endoscope.
(1)
Visually inspect the boot and the insertion portion near the boot for bends, twists, swelling or
other irregularities.
(4)
(2)
(2)
Visually and manually check the insertion portion (distal end, bending section and insertion
tube) for abnormalities such as indentations, bumps, peeling, sharp edges or protrusions, etc.
In addition, confirm that the insertion tube is not abnormally rigid.
(3)
Visually and manually check the bending section sheath and adhesives at each end of the
bending section for abnormalities such as missing epoxy, roughness, pitting or flaking.
ED580XT_E2-30_897N200877B.indb 6
2018/09/20 15:50:20
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...