897N200877B
5-10
Chapter 5 How to Use
5.2.6 Operating Suction Valve
WARNING
• Set the suction pressure between 40 and 53 kPa. If the suction pressure is too high,
patient debris or fluids may leak or splash from the forceps valve, posing infection control
risks to patient or operator.
• Do not quickly release one’s finger from the suction valve during aspiration. Doing so may
cause a splattering of body fluids from the suction valve and increase a risk of infection.
• The lid of the forceps valve must be closed when using the endoscope. Not doing so may
cause leak of body fluids and increase a risk of infection.
• Avoid aspirating solid materials or thick fluids. If the suction valve does not return to its
original position, stop aspiration immediately and slowly withdraw the endoscope. If any
solid materials or thick fluids adhere to or clog the suction valve, suction may not stop,
causing damage to mucous membrane.
CAUTION
• Set the suction pressure between 40 and 53 kPa. If the suction pressure is too high,
the endoscope may adhere to mucous membrane, resulting in damage to the mucous
membrane.
• The lid of the forceps valve must be closed when using the endoscope. Not doing so can
reduce the efficacy of the endoscope’s suction system, making it impossible to perform
aspiration.
(1)
(1)
Press the suction valve to aspirate fluids in the body cavity or patient materials adhering to the
distal end from the instrument channel outlet.
ED580XT_E2-30_897N200877B.indb 10
2018/09/20 15:50:24
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...