897N200877B
4-9
Chapter 4
Prepara
tion and Inspectio
n
<When the up/down and left/right angulation knobs are in the neutral positions>
(4)
When the up/down and left/right angulation knobs are turned to their respective neutral
positions, visually check that the bending section returns smoothly to an approximately straight
condition.
♦
Inspecting Up/Down Angulation
U
D
(1)
Move the up/down angulation lock all the way in the opposite direction of F (Free) to lock the
up/down angulation knob.
Note
When the up/down or left/right angulation lock is moved in the opposite direction of F
(Free), the rotation of the up/down or left/right angulation knob becomes harder, and
even when fully locked the angulation knob and the distal bending section still can be
moved.
(2)
Turn the up/down angulation knob in the directions of U and D until it stops.
Confirm that the angle of the bending section is roughly stabilized when the up/down
angulation knob is released.
(3)
Move the up/down angulation lock all the way in the direction of F (Free) to unlock the up/down
angulation knob.
Confirm that the bending section moves back toward its neutral position when the up/down
angulation knob is released.
4.3 Inspecting Endoscope
Insertion Portion
Distal End
Bending Section
LG Connector
Control Portion
Forceps Elevator
ED580XT_E2-30_897N200877B.indb 9
2018/09/20 15:50:21
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...