897N200877B
4-8
Chapter 4 Preparation and Inspection
4.3.4 Inspecting Bending Section
WARNING
• Turn the up/down and left/right angulation knobs slowly in each direction until they stop.
Repeat this operation several times to confirm that the bending section angulates smoothly
and correctly. If the endoscope with an abnormal angulation knob is used, the bending
section does not return to its neutral position, causing patient injury.
CAUTION
• Do not forcibly turn the angulation knob further after turning the knob until it stops. If the
angulation knob is forcibly turned, it may cause malfunction of the endoscope.
♦
Inspection for Smooth Operation
(1)
Straighten the bending section.
(2)
Turn the up/down and left/right angulation locks in the direction of F (Free) until they stop to
unlock the up/down and left/right angulation knobs and confirm that the up/down and left/right
angulation knobs move smoothly.
(3)
Slowly turn the up/down angulation knob in the directions of U and D and the left/right
angulation knob in the directions of L and R. Ensure the distal tip moves to the desired
direction, and then return them to their neutral position. Repeat this operation several times
to confirm that the bending section angulates and returns to its neutral position smoothly and
correctly.
ED580XT_E2-30_897N200877B.indb 8
2018/09/20 15:50:21
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...