897N200877B
4-16
Chapter 4 Preparation and Inspection
4.4.3 Air/Water Valve
Skirt
Check-Valve
Horizontal Hole
Rubber Seal
Spring
Hole
Air/Water Valve (AW-603)
WARNING
• Use a properly reprocessed air/water valve. Use of an improperly reprocessed air/water
valve could be a source of infection.
• Use the air/water channel cleaning adapter only for pre-cleaning of the air/water channel.
If it is used during a procedure, continuous air supply may occur and cause patient injury.
CAUTION
• Do not use any lubricants (except sterile water) to the air/water valve. It may impair the
functionality of the valve or may clog the channel, diminishing functionality of air/water
supply.
<Preparation>
The air/water valve is not reprocessed before shipping from FUJIFILM. Reprocess it according
to the instructions given in the Reprocessing Manual before using it in a procedure.
Reprocessing Manual
<Inspection>
(1)
Visually check that the hole of the air/water valve is not blocked.
(2)
Visually check that the air/water valve is not deformed or cracked.
Note
The air/water valve is a consumable item. If any abnormality is found, use a new
reprocessed valve.
(3)
Visually check the air/water valve for excessive scratching or tears in the valve’s seals.
ED580XT_E2-30_897N200877B.indb 16
2018/09/20 15:50:22
Summary of Contents for ED-580XT
Page 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Page 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Page 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Page 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Page 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Page 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...