4.3.12
PLC
Software
Update
The
following
is
the
procedure
for
updating
the
software
in
the
P2131
Automated
Field
Steam
Sterilizer
using
the
Micro
SD
card
that
is
shipped
with
each
sterilizer.
NOTE:
This
procedure
should
only
be
performed
by
a
qualified
Maintenance
Technician
1.
The
updated
software
file
will
be
delivered
via
email.
2.
Using
the
Micro
SD
card
and
adapter
supplied
with
each
sterilizer,
save
the
file
onto
the
Micro
SD
card
with
your
computer.
3.
Ensure
the
sterilizer
is
OFF.
4.
Insert
the
Micro
SD
card
(without
the
adapter)
into
the
extender
cable
and
insert
the
other
end
of
the
cable
into
the
top
of
the
Programmable
Logic
Controller
(PLC).
The
exposed
pins
should
face
toward
the
front
of
the
sterilizer.
5.
Note
the
firmware
version
number
by
turning
the
sterilizer
ON
and
watching
the
first
screen
that
appears.
The
firmware
version
number
will
display
for
only
2
seconds.
6.
From
the
CONFIRM
SETUP
screen,
press
the
F4
function
key.
(F4
is
only
accessible
from
this
screen.)
7.
The
UPDATE
SOFTWARE
screen
will
display
and
the
PLC
will
initialize
the
Micro
SD
card.
This
may
take
a
few
minutes
after
which
a
list
of
files
will
appear.
8.
Use
the
up
and
down
arrow
keys
on
the
right
‐
hand
side
of
the
screen
to
select
the
new
software
file.
(Example:
V1285
OR
V1043)
9.
Choose
OK
when
you
are
asked
if
you
are
sure
you
want
to
load
the
application.
10.
A
moving
clock
and
progress
bars
will
appear
while
the
software
is
loaded.
This
may
take
a
few
minutes.
11.
Next,
you
will
see
LOAD
PAGE
COMPLETE
and
you
will
be
asked
if
you
would
like
to
place
the
PLC
in
RUN
mode.
Choose
OK.
(Choosing
CANCEL
will
place
the
PLC
in
IDLE
mode
and
the
software
will
not
operate.)
12.
The
software
will
restart
and
the
CONFIRM
SETUP
screen
will
appear.
13.
Turn
the
sterilizer
OFF
and
remove
the
Micro
SD
card
and
extender
cable.
14.
The
new
software
is
now
installed
and
the
sterilizer
is
ready
to
operate.
4.3.13
PLC
Firmware
Update
The
following
is
the
procedure
for
updating
the
firmware
in
the
P2131
Automated
Field
Steam
Sterilizer
using
the
P2131
FIRMWARE
UPDATE
KIT
from
Fort
Defiance
Industries.
NOTE:
This
procedure
should
only
be
performed
by
a
qualified
Maintenance
Technician
92
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...