
Choosing
to
“RESET”
will
either
(1)
switch
the
display
to
the
“Log”
screen
or
(2)
repeat
the
same
alarm
if
the
condition
is
still
present.
The
controller
will
not
allow
the
sterilizer
to
continue
until
the
alarm
has
been
reset.
For
example,
if
the
alarm
was
“02
‐
Low
Water
Level,”
water
will
need
to
be
added
to
the
jacket
before
the
alarm
will
reset.
The
controller
will
then
display
the
“Log”
screen
with
the
details
of
the
alarm
for
recordkeeping.
3.3.6
MAINTENANCE
REMINDER:
Per
FDA
regulations,
every
100
cycles
or
14
calendar
days
,
the
water
in
the
entire
system
must
be
drained
and
refilled.
After
every
100
cycles,
a
“Maintenance
Reminder”
screen
is
shown
to
remind
the
operator
that
the
water
in
the
SMU
jacket,
the
WRS,
and
the
SWS
must
be
drained
and
replenished
with
fresh
potable
water.
If
needed,
flush
the
SWS
until
clear
(2
‐
minute
flush).
This
reminder
also
serves
to
help
encourage
maintenance
technicians
to
check
and
follow
the
recommended
maintenance
schedule
(Reference
Section
4).
3.3.7
STATUS:
The
“STATUS”
screen
can
be
accessed
(except
when
a
cycle
is
running)
in
order
to
look
at
data
such
as
jacket
pressure,
chamber
pressure,
chamber
condensate
temperature
(both
RTDs),
whether
the
door
switch
is
showing
open
or
closed,
if
the
water
level
is
OK
or
low,
and
how
many
total
cycles
the
unit
has
run.
This
screen
can
be
accessed
by
selecting
the
“STATUS”
button
from
the
“Main
Menu”
screen.
47
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...