
SMU
–
Water
and
Steam
Related
Situations
What
do
you
do
if
…
Response
You
hear
or
see
a
steam
leak
somewhere
other
than
the
door?
If
it
is
safe
to
do
so,
turn
the
control
power
switch
OFF
and
allow
the
sterilizer
to
cool
and
depressurize.
Contact
a
qualified
maintenance
technician.
You
hear
or
see
a
steam
leak
at
the
door?
If
it
is
safe
to
do
so,
wait
for
the
cycle
to
finish
and
then
inspect
the
door
frame
and
door
gasket
for
any
debris
that
could
prevent
a
good
seal.
Closely
inspect
the
gasket
for
any
damage
such
as
nicks
or
cuts.
Replace
it
if
needed.
If
there
is
no
damage,
then
tighten
the
door
a
little
tighter
on
the
next
cycle.
You
get
a
low
water
alarm?
Add
water
to
the
SMU
from
the
WRS,
close
and
tighten
the
chamber
door,
press
“RESET,”
and
restart
the
cycle.
Also,
recheck
level
in
WRS
after
pumping
and
refill
using
the
SWS
if
needed.
It
is
best
to
run
the
WRS
at
the
HIGH
water
mark
(when
the
jacket
is
at
the
HIGH
mark).
The
jacket
is
overfilled
with
water?
Turn
the
sterilizer
OFF
and
let
it
cool
down
and
depressurize.
With
zero
pressure
on
the
jacket,
attach
the
high
‐
temperature
hose
and
drain
enough
water
out
of
the
jacket
to
bring
the
level
down
to
the
HIGH
mark
on
the
sight
gauge.
Close
the
drain
valve
and
turn
the
sterilizer
back
ON.
The
jacket
fills
with
water,
but
VERY
slowly?
Contact
a
qualified
maintenance
technician
for
further
diagnosis
and
repair
on
the
WRS
pump.
The
jacket
does
not
refill
when
switch
on
WRS
depressed?
Verify
the
water
level
in
the
WRS,
adding
water
through
the
SWS
if
necessary.
Ensure
that
the
power
cord
is
plugged
into
the
WRS
and
that
the
power
switch
is
in
the
ON
position.
If
the
problem
continues,
contact
a
qualified
maintenance
technician
for
further
diagnosis
and
repair.
The
water
level
in
jacket
drops
when
sitting
idle
and
pressurized?
The
check
valve
at
the
back
of
SMU
on
recovered
water
line
from
WRS
could
be
leaking.
Contact
a
qualified
maintenance
technician
to
replace
the
check
valve.
The
SMU
seems
to
be
using
more
water
than
normal
during
each
cycle?
The
check
valve
at
the
back
of
SMU
on
the
recovered
water
line
from
WRS
could
be
leaking.
Additionally,
the
steam
trap
could
be
hung
open
or
damaged.
Contact
a
qualified
maintenance
technician
to
inspect/replace
the
check
valve
and/or
the
steam
trap.
The
Pressure
Safety
Valve
opens
and
releases
steam?
Contact
a
qualified
maintenance
technician.
If
it
is
safe,
turn
the
sterilizer
OFF
and
disconnect
the
incoming
power.
There
is
a
lot
of
water
around
the
sterilizer?
The
source
of
the
water
should
be
identified
immediately.
Inspect
the
hose
connections
between
the
SMU
and
the
WRS.
Use
a
flashlight
to
inspect
and
locate
the
leak.
If
needed,
de
‐
energize
and
de
‐
pressure
the
sterilizer,
remove
the
side
panels
and
inspect
all
fittings
closely.
White/Tan
residue
in
chamber
Dirty
and/or
hard
water
used.
SWS
not
used.
Possible
cracked
heater
element.
Inspect
heaters
and
replace
if
needed.
(See
Section
4.3.10)
58
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...