3.3.8
PROGRAMMABLE
CYCLES:
Programmable
cycles
allow
the
user
to
run
cycles
with
exposure
times
that
can
range
from
15
minutes
to
40
minutes
in
increments
of
5
minutes.
Dry
times
can
range
from
0
to
60
minutes,
also
in
5
‐
minute
increments.
1.
To
run
a
programmable
cycle,
press
“CUSTOM”
from
the
“Main
Menu”.
2.
After
carefully
reading
the
WARNING
,
press
“Accept”
to
continue.
3.
Press
the
arrow
keys
to
select
the
desired
exposure
and
dry
times,
and
then
press
“ACCEPT”.
WARNING:
Only
the
standard
AAMI
cycles
listed
in
Section
2.3
have
been
cleared
by
the
FDA.
Programmable
cycles
have
not
been
validated.
The
FDA
highly
recommends
that
programmable
cycles
be
used
for
research
purposes.
The
user
accepts
all
risks
associated
with
sterilizing
any
load
using
a
programmable
cycle.
48