4.3.4
Replacing
the
SWS
Cartridge
1.
Remove
inlet
and
outlet
hoses
from
the
blue
housing.
2.
For
S/N’s
AFS
‐
0001
through
AFS
‐
0255,
remove
the
bungee
cords
and
slide
the
blue
housing
out
from
the
SWS
frame.
For
S/N’s
after
AFS
‐
0255,
remove
the
bracket
that
secures
the
blue
housing
by
loosening
the
two
panel
fasteners
and
slide
the
blue
housing
out
from
the
SWS
frame.
3.
Use
the
supplied
spanner
wrench
stored
inside
the
SWS
frame
to
remove
the
blue
housing
from
the
black
top.
4.
Slide
the
old
cartridge
out
and
discard.
5.
Slide
the
new
cartridge
in
place,
ensuring
the
gasketed
side
is
up
towards
the
opening
of
the
housing.
6.
Verify
large
Blue
Housing
O
‐
ring
is
in
place
and
is
lubricated
with
silicone
grease.
7.
Follow
steps
1
‐
3
in
reverse
order
to
return
the
SWS
to
operation.
4.3.5
Replacing
the
HEPA
Filter
1.
Turn
OFF
the
SMU,
disconnect
power,
and
ensure
pressure
in
the
chamber
and
jacket
is
at
0
psig.
2.
Remove
the
left
side
panel
of
the
SMU.
3.
Using
an
11/16”
wrench
to
counter
‐
torque
the
fitting
near
solenoid
valve,
use
an
11/16”
wrench
to
loosen
the
fitting
nut
at
the
filter.
4.
Install
the
new
HEPA
filter
by
following
steps
1
‐
3
in
reverse
order.
4.3.6
Replacing
the
Steam
Trap
1.
Turn
off
the
SMU,
disconnect
power,
and
ensure
pressure
in
the
chamber
and
jacket
is
at
0
psig.
2.
Disconnect
all
three
hoses
from
the
WRS.
Counter
Torque
fitting
Loosen
Nut
Blue
Housing
O
‐
ring
Gasket
Side
Up
86
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...