Function(s)
of
Control
or
Indicator
(Rear
View
of
SMU)
ITEM
CONTROL/INDICATOR
FUNCTION
1.
Water
Level
Sight
Gauge
Indicates
water
level
inside
jacket.
Fill
to
HIGH
mark.
A
low
water
level
switch
(separate
from
sight
gauge
and
not
shown
in
picture)
will
create
an
alarm,
abort
the
cycle,
and
de
‐
energize
the
heaters.
2.
Pressure
Safety
Valve
(PSV)
Relieves
overpressure
@
40
psig
to
prevent
potential
rupture
of
jacket/chamber
3.
Jacket
Vent
Valve
Manual
valve
that
opens
the
jacket
to
atmosphere.
This
valve
must
be
closed
during
normal
operation.
Best
practice
includes
NOT
opening
this
valve
after
shutdown
to
vent
steam,
but
allow
SMU
to
cool
and
steam
to
condense
(this
will
pull
a
vacuum
in
jacket,
which
is
OK).
4.
Chamber
Vent
Valve
Manual
valve
that
opens
the
chamber
to
atmosphere.
This
valve
must
be
closed
during
normal
operation.
This
valve
is
used
to
relieve
pressure
or
vacuum
(e.g.,
if
vacuum
develops
in
chamber
from
accidently
leaving
door
closed
during
cool
down).
5.
Recovered
Water
Hose
Connection
(Blue
‐
Tagged)
This
hose
supplies
water
to
the
jacket
from
the
WRS.
6.
Vacuum/Exhaust
Hose
Connection
(White
‐
Tagged)
This
hose
exhausts
steam/air
and
allows
the
WRS
to
create
the
vacuum
necessary
for
sterilizer
operation.
7.
Jacket
and
Chamber
Manual
Vent
Valve
Outlets
Outlet
points
for
the
Jacket
and
Chamber
vent
valves.
WARNING
:
A
small
amount
of
hot
condensate
will
collect
above
these
valves
during
operation.
Slowly
crack
the
valve
to
allow
any
hot
condensate
to
empty
out,
and
keep
hands
away
from
this
outlet
point.
8.
Jacket
Drain
Hose
Connection
Hose
connection
draining
hot
water
from
jacket.
ONLY
use
the
high
temp
hose
provided
to
drain
jacket.
Ensure
end
of
hose
is
secured
to
prevent
hose
whip
in
the
unlikely
event
of
pressurized
steam
exhausting
from
this
hose.
WARNING
:
A
standard
garden
hose
could
rupture,
resulting
in
potential
burns.
9.
Jacket
Drain
Valve
Manual
valve
that
opens
the
jacket
to
atmosphere
and
allows
the
jacket
to
drain.
This
valve
must
be
closed
during
normal
operation.
Do
not
open
this
valve
with
pressure
on
jacket.
10.
Heater
Flange
Drain
Caps
S/N’s
AFS
‐
0001
through
AFS
‐
0236:
Used
to
drain
heater
flange
tees
for
transport
and/or
storage.
11.
Multi
‐
tool
Used
during
the
setup
of
the
sterilizer
to
tighten
hoses
and
remove
panels.
12.
Tee
Cap
Retainer
S/N’s
AFS
‐
0001
through
AFS
‐
0236:
Used
to
store
Heater
Flange
Drain
Caps
during
Transport
and/or
Storage
of
the
Sterilizer.
13.
Back
Lower
Panel
This
panel
prevents
the
user
from
accessing
underneath
the
Chamber.
14.
Heater
Assembly
Cover
This
cover
protects
the
user
from
any
potential
electrical
shock
hazard.
33
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...