ALARM
CODE
POSSIBLE
CAUSES
CORRECTIVE
ACTION
08
Chamber
Pressure
Low
Gasket
leak
at
door
Manual
chamber
vent
valve
leaking
Faulty
steam
trap
Vacuum/Exhaust
valve
leaking
Sterilize
valve
failure
–
not
opening
properly
Condensate
drain
screen
plugged
Inspect
for
leak
with
pressure
and
identify
location.
Inspect
gasket
for
debris
or
cut.
Replace
as
needed.
Contact
a
qualified
maintenance
technician
to
inspect
vent
valve
and
replace
as
needed.
Ensure
trap
is
working
properly.
If
steam
trap
is
failed
closed,
condensate
will
pool
in
chamber.
If
steam
trap
fails
open,
steam
will
flow
to
WRS
water
tank
and
make
loud
popping
noise
or
use
more
water
than
normal.
Replace
as
needed.
Contact
a
qualified
maintenance
technician
to
inspect
outlet
of
valve
for
leaking.
Contact
a
qualified
maintenance
technician
to
inspect
solenoid
valve
and
replace
if
needed.
Remove,
clean,
and
replace
screen.
NOTE:
Also
reference
causes
of
Alarm
Code
05
Long
Exposure
Time.
64
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...