
ALARM
CODE
POSSIBLE
CAUSES
CORRECTIVE
ACTION
09
Jacket
Pressure
High
Air
compressed
in
jacket
during
water
fill
RTD
or
RTD
transmitter
failure
SMU
tilt
is
not
steep
enough
Condensate
drain
screen
plugged
Solid
state
heater
relays
leaking
current
Jacket
pressure
transmitter
failure
Turn
OFF
sterilizer
and
allow
jacket
to
cool
and
depressurize.
Open
jacket
vent
valve
to
bring
pressure
to
zero.
Close
jacket
vent
valve.
Turn
sterilizer
back
ON.
Contact
a
qualified
maintenance
technician
to
inspect
RTD
and
RTD
transmitter
for
physical
damage,
loose
wires,
broken
communication
cable
to
PLC.
Compare
temperature
versus
pressure
on
STATUS
screen.
Replace
as
needed.
Shim
the
rear
of
the
sterilizer
higher
to
improve
condensate
flow
to
the
chamber
drain.
Remove
and
clean
screen.
Replace
after
cleaning.
Contact
a
qualified
maintenance
technician
to
test
relay
for
leakage
and
replace
as
needed.
Open
Chamber
Door
and
compare
digital
pressure
readout
on
STATUS
screen
on
PLC
to
analog
gauge
on
front
panel.
Jacket
Pressure
reading
≈
28
psig
and
Chamber
Pressure
Reading
≈
0
psig.
10
Long
Vent
Time
Barometric
Pressure
Not
Zeroed
Door
not
tightened
during
manual
VENT
Vacuum
break
valve
failure
(not
opening
far
enough)
Vacuum/exhaust
valve
failure
(not
opening
far
enough)
if
on
IUSS
cycle
On
“CONFIRM
SETUP”
screen,
zero
barometric
pressure.
This
is
only
required
during
setup
at
a
new
location.
Tighten
door.
Contact
a
qualified
maintenance
technician
to
inspect
and
test
both
valves.
Replace
as
needed
(Note
flow
direction.
Arrows
on
valve
should
point
toward
each
other
(
).
Open
chamber
vent
valve
to
access
chamber.
Contact
a
qualified
maintenance
technician
to
inspect
and
test
valve.
Replace
as
needed.
Open
chamber
vent
valve.
65
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...