Biological Indicators:
The reusable test packs have been validated with biological indicators commonly used to monitor steam sterilizers. Table 1 lists the
validated biological indicators.
Table 1: Validated Biological Indicators.
Biological Indicator
Product Number
Manufacturer
3M Attest Rapid Readout Biological Indicator
1292
3M Company
3m Attest Super Rapid Readout Biological Indicator
1492V
3M Company
Smart-Read EZTest Self Contained Biological Indicators for Steam
SEZS/5 and SEZS/6
MesaLabs, Inc.
VERIFY STEAM Integrating Indicators:
The VERIFY STEAM Integrating Indicator is used on its own or in conjunction with a biological indicator for routine monitoring
and release of steam sterilized loads.
INTEGRATING INDICATOR i5
VERIFY™
STEAM
REJECT (NO)
ACCEPT (OK)
DARK BAR MUST ENTER ACCEPT WINDOW
LOT 615-A EXP. 20200601
During the test cycle, the steam sensitive chemical located within the strip will wick across the “REJECT” viewing window and enter
into the “ACCEPT” viewing window. Sufficient steam has entered the reusable test pack when any of the dark bar is visible within the
“ACCEPT” Viewing window.
INTEGRATING INDICATOR i5
VERIFY™
STEAM
REJECT (NO)
ACCEPT (OK)
DARK BAR MUST ENTER ACCEPT WINDOW
LOT 615-A EXP. 20200601
INTEGRATING INDICATOR i5
VERIFY™
STEAM
REJECT (NO)
ACCEPT (OK)
DARK BAR MUST ENTER ACCEPT WINDOW
LOT 615-A EXP. 20200601
ACCEPT
REJECT
138
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...