ALARM
CODE
POSSIBLE
CAUSES
CORRECTIVE
ACTION
21
Temperature
Sensor
Large
RTD
deviation,
RTD
or
RTD
transmitter
failure
Contact
a
qualified
maintenance
technician
to
inspect
RTD
and
RTD
transmitter
for
physical
damage,
loose
wires,
broken
communication
cable
to
PLC.
Compare
RTD
‐
1
versus
RTD
‐
2
on
“STATUS”
screen.
Replace
as
needed.
22
High
Exposure
Temperature
RTD
or
RTD
transmitter
failure
Faulty
steam
trap
Condensate
drain
screen
plugged
Sterilizer
not
tilted
correctly
Solid
state
heater
relays
leaking
current
Contact
a
qualified
maintenance
technician
to
inspect
RTD
and
RTD
transmitter
for
physical
damage,
loose
wires,
broken
communication
cable
to
PLC.
Compare
temperature
versus
pressure
on
STATUS
SCREEN.
Replace
as
needed.
Ensure
trap
is
working
properly.
If
steam
trap
is
failed
closed,
condensate
will
pool
in
chamber.
If
steam
trap
fails
open,
steam
will
flow
to
WRS
water
tank
and
make
loud
popping
noise
or
use
more
water
than
normal.
Replace
as
needed.
Remove,
clean,
and
replace
screen.
Verify
correct
tilt.
Water
should
drain
easily
to
front
chamber
drain.
Contact
a
qualified
maintenance
technician
to
test
relay
for
leakage
and
replace
as
needed.
23
Jacket
Sensor
Jacket
pressure
transmitter
failure
Contact
a
qualified
maintenance
technician
to
inspect
and
replace
transmitter.
68
Summary of Contents for P2131
Page 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Page 130: ...SECTION A APPENDIX 124 ...
Page 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Page 135: ...A 2 PLC Data Sheet 129 ...
Page 141: ...A 3 VERIFY Product Information 135 ...
Page 147: ...A 4 WRS Pump Information 141 ...
Page 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...