MP101M
Environnement
S.A
5–5
SEPTEMBER 2013
DISPLAY CAUSE
POSSIBLE
ACTIONS
Threshold D.
–
The periodic or cyclic
measurement exceeds
the superior allowed
value (1000 µg/m
3
or
1000 mg/m
3
).
–
Change the measurement units: replace
µg/m
3
by mg/m
3
in the
« CONFIGURATION
Offsets and units »
screen.
–
Check the source positioning in front of the
paper. If it is not correctly positioned, check
correct operation of the positioning motor.
–
If the motor does not work, check the + 15V
and - 15V voltages.
Clogging
–
The system remained
more than 360 s (6 min)
running in regulation
valve control mode and
:
1013
750
1
A
P
P
.
–
Check the pressures: Upstream pressure,
Downstream pressure, Atmospheric
pressure.
–
Check the correct operation of the valve
motor control (refer to the screen
« TESTS
Mother board Inlets/outlets »
).
–
Check the valve motor.
–
Check the valve.
–
Check the fluid circuit.
Flow rate
–
The flow rate exceeds
the allowed interval
centered around the 1
m
3
/h control.
–
Check the correct operation of motor valve
control (refer to the screen
« TESTS
Mother board Inputs/Outputs »
).
–
Check the valve motor.
–
Check the valve.
–
Test for leaks in the fluid circuit.
R.A.N. counting
–
The analyzer detects an
abnormally high rate of
natural radioactivity.
–
Inform the appropriate authorities as soon as
possible.
Paper torn up
–
Counting value provided
by the Geiger-Müller
> 8500 counts/s.
–
Check the position of the filter paper ribbon.
Reposition it again.
–
Carry out a contamination test.
–
Check the source positioning in front of the
paper. If the source is not correctly
positioned, check correct operation of the
positioning motor. If the motor does not work:
check the + 15V and - 15V voltages.
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...