It gives the summary of the last calibration with :
The number of calibration cycles performed, the counting duration for each cycle, the value of the
reference gauge, stated into µg/cm
2
.
The average values of blanks and calibrations with the average of their respective measurement
temperatures.
The « K final » or final span factor obtained at the end of the calibration and stored into memory.
In this screen, pressing down the [Blanks] key enables the user to display the screen shown below:
This screen displays:
The detail of blank measurements with the value of filter temperature when the blank
measurement is performed,
The average values of blanks and temperature, as displayed in the previous screen.