Duplication prohibited
MP101M
Environnement
S.A
1–11
SEPTEMBER 2013
–
Details of the Flow regulation part (
Figure 1–8
)
The main components are:
one motor-driven valve (1),
a block with a 2.8 mm diameter flat orifice (3).
The hydrostatic pressure drop observed on passage at the throttle in the fluid circuit is used to
deduce the velocity of flow and the flow rate.
three pressure sensors upstream and downstream of the flat orifice : P1 (4), P2 (2) and
atmospheric P not linked to the fluid circuit (5).
(1) motor-driven valve, (2) P2 downstream pressure sensor, (3) flat orifice, (4) P1 upstream pressure
sensor, (5) atmospheric pressure sensor (P. atm.).
Figure 1–8 – Details of the flow regulation part
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...