3.3.4.8 CONFIGURATION
Communication
UDP server
UDP stands for User Datagram Protocol. The analyzer network communication runs only under this
protocol.
UDP applications use datagram sockets to establish host-to-host communications.
These sockets bind the application to service ports that function as the endpoints of the data
transmission. A port is a software structure that is identified by a port number. Two different ports can
be bind to the analyzer: Port A and Port B. They are identified by two different corresponding port
numbers (8003 and 8001 for example).
Port
fields : the numbers of Port A and Port B of the UDP server are programmable from 1000 to 9999
Address
fields
:
they are programmable with 4 characters, the analyzer’s name is used by default.
Protocole
fields: the user can choose different types of protocols: Mode4, JBUS, and PRN. In the last
case (PRN), the IP address of the computer has to be entered.
3.3.4.9 CONFIGURATION
Factory setting
When this item is selected, pressing down the [
] key displays the screen shown below:
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...