Duplication prohibited
MP101M
Environnement
S.A
1–15
SEPTEMBER 2013
Output signals
: up to 8 analog outputs (programmable full scale )
0-1 V, 0-10 V, 0-20 mA, or 4-20 mA with
galvanic isolator on option
(depending on the number of ESTEL boards option
present)
Power supply
: 230 V-50 Hz (115 V-60 Hz on request) + ground
Consumption
: Without pump: 350 mA
With pump: 1500 mA
Maximum active power
: 200 VA
consumed by pump
Power of unit
: 300 VA
Operating temperature
: + 10 °C to + 40 °C
Alarm checks
: – Permanent
–
Detection and indication of malfunctions: temperature,
pressure, flow rates, electric parameters.
Tests and diagnostics for maintenance : Selection on keyboard and display of all parameters.
Back-up saving time for data stored
in FLASH
:
unlimited
and of real-time clock:
:
2 years by built-in battery
1.2.2.
OPERATING CHARACTERISTICS
Refer to Chapter 0, Safety Guidelines section, pages 0-10 to 0-18
1.2.3. STORAGE
CHARACTERISTICS
–
Temperature: – 10 °C to 60 °C.
–
This model uses a rotary source holder. The Carbon 14 source can thus take two positions:
storage and emission.