MAINTENANCE SHEET
Monitor serial No.::
OPERATION SHEET: 4.3.7
Scope : Flow rate and leak rate tests
PAGE : 1/2
Periodicity : 3 months
Dates
Procedure :
Carry out a flow rate test and a leak rate test after each replacement of filter ribbon. If necessary,
repeat a flow rate calibration.
1/ FLOW RATE TEST :
–
Remove the MP101M from its sampling line.
–
Connect a flow meter (equipped with the fitting P02-1571) on the sample inlet.
–
Go to the «
CALIBRATION
Flow rate ».
screen. Press down the [>>] key, then the [Auto] key,
then the [Start] key. The pump -up and the flow rate measured by the analyzer is displayed on
the screen under the reference flow rate.
–
Press down the [Details] key and connect the reference flow meter on the retractable fitting at
sample inlet.
–
Check that the flow rates respectively measured by the flowmeter (« Span ») and by the
analyzer (« Measurement ») are consistent:
–
If the variance between these two measurements is lower than 1.5 L/min, carry out a flow rate
calibration (cf. paragraph 3.3.3.2).*
Remark : This test can also be carried out upstream the sampling line, with the reference flow meter
placed instead of the sampling head. In this case, two people will be necessary to carry out this test
easily.
2/ TEST FOR LEAKS :
–
Go to the
« TESTS
Mother board Inputs - outputs »
screen.
–
Go
to
Nozzle
then press down the [Test] key then press down the [ s ] key under S
ampling
to
set in action the motor of the source holder
The source is put back in its place and the filter
holder is placed into sampling position. The motor stops after five seconds.
–
Go
to
Plate
then press down the [Test] key, then press down the [ s ] under
Mov. Down
to set in
action the plate motor
The plate goes down. The motor stops after five seconds.
–
Go
to
Filter
then press down the [Test] key then press down [ s ] to set in action the motor for
moving on paper
The paper moves on. The motor stops after 7 seconds.
–
Go to the
ON/OFF
field in the line of the
Pump
field. Press down the [
] key to modify the
parameter into ON and set the pump into action
the pump starts-up,
the analyzer is in
sampling
.
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...