Duplication prohibited
MP101M
Environnement
S.A
4–17
SEPTEMBER 2013
Monitor serial No.:
OPERATION SHEET : 4.3.5
Scope : Replacement of filter ribbon
PAGE : 3/3
Periodicity : from 1 to 3 years
Periodicity : W
HEN THE
ALARM
«
PAPER
END
»
IS DISPLAYED
(1
TO
3
YEARS ACCORDING TO
PROGRAMMING OF THE CYCLE DURATION
).
Dates
11/ It is now necessary to reset the counter of paper remaining capacity: go to the «
CONFIGURATION => Measurement mode
» screen and press down the [>>] key. Using the [
]
key, go to the
Reel
line on the
ON/OFF
field and press down the [
] key to modify the parameter
OFF
into
ON
. The
spots
number is reset to 1200.
12/ After each replacement of ribbon filter, it is strongly advised to :
–
Check the pressure on the filter ribbon.
–
Test for leaks in the sampling system.
–
Calibrate the analyzer (Flow rate and Beta gauge).
–
Tools required
Roll of filter paper RF100 ref. M04-370-392
Adhesive
tape