MP101M
Duplication prohibited
SEPTEMBER 2013
3–56
(3)
When the cursor indicates « Start », press down the [Start] key to launch a mass test.
The screen shown above is replaced by this one :
(4)
The analyzer carries out an initialization by moving ahead a new part of filter paper towards the
measurement position. Then, it controls the pump switching on which will operate during 120 seconds
in order to dry the filter paper. After what, it carries out a blank measurement.
When the blank measurement is finished, the cursor moves to « Insert reference gauge » position and
this message blinks.
The user has to place the reference gauge, guided by the two pins, until hearing a click sound when it
arrives in limit stop (see Figure 3–4). Once the reference gauge is installed, the user has to validate
this operation by pressing down the F3 key.