MAINTENANCE SHEET
Monitor serial No.:
OPERATION SHEET: 4.3.6
Scope : Verification of pressure exerted on the filter
PAGE : 1/2
Periodicity : 1 year
Dates
Procedure :
1/ Stop any measurement in progress, if necessary.
2/ Go to the «
TESTS
Mother board Inputs - Outputs »
screen and make the plate moves down :
go to the
Plate
field then press down the [Test] key, then the [ s ] key under
Mov. down
.
3/ Open the door of the analyzer using the knurled button located on the left.
4/ Raise the disengageable pinch roller, as shown in the just below picture :
5/ Insert a "smooth" paper between the filter ribbon and the driving capstan (the right one). Only the
capstan rotation must cause the moving on of the ribbon filter, as shown in the just below picture.
6/ In the «
TESTS
Mother board inputs - outputs »
screen
,
make the paper moves on
continuously by pressing down the [ ... ] key under
Test
after having selecting the
Filter
line with the
F6 key.
7/ Keep in position the paper (inserted between the ribbon and the capstan) to prevent its driving on
by the capstan.
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...