Duplication prohibited
ESTEL Board
Environnement
S.A
JUNE 2009
13
1.5
INSTALLATION AND REPLACEMENT OF ESTEL BOARD
User’s safety recommendation : Switch off the analyzer and unplug the main cable before any
maintenance work of the analyzer.
Analyzer’s safety recommendation: Respect the connection of ESTEL board / Module board at
J20 when reassembling.
Use the steps illustrated below for installation or replacement of the ESTEL board.
1.5.1
Switch off analyzer.
1.5.2
Unplug main cable.
1.5.3
Take off cover.
(2) Unscrew screws on lateral sides.
(1) Unscrew screws on rear panel of analyzer.
(3) Lift up cover.
(4) Remove cover by pulling it backwards.