![Environnement MP101M Technical Manual Download Page 150](http://html1.mh-extra.com/html/environnement/mp101m/mp101m_technical-manual_2416152150.webp)
Environnement
S.A
MP101M
Duplication prohibited
SEPTEMBER 2013
4–28
MAINTENANCE SHEET
Monitor serial No.::
OPERATION SHEET: 4.3.12
Scope : Verification / replacement of temperature and
relative humidity sensors.
PAGE : 1/1
Periodicity : 6 months
Dates
Procedure :
1/ Go to the
« I2C CARD(S)
RST board »
screen.
2/ Check the values of the atmospheric temperature and relative humidity using reference sensors.
If relative humidity is lower than 60%, check that Head temperature = Atmos. Temperature.
If relative humidity is higher than 60%, check that Head temperature = Atmos. Tempe 5°C.
3/ If necessary, adjust the A and B linearization coefficients of the three sensors.
4/ In case of replacement of the meteo sensor ( T°C and RH sensors) :
Refer to the calibration report of the relative humidity (RH) sensor provided :
Number sensor
Channel
...........................................
Pente (format xx,xxx)
Slope
S = .....................................
mV / %RH
Offset à zéro (format xx,xxx)
Zero offset
Z = .....................................
V
And on the
Relative humidity
field of the «
I2C CARD(S)
RST board »
screen, replace the values
of the A and B calibration coefficients by :
A = 1 / S
B = - Z x 1000 / S
–
Tools required
Reference
gauge
–
Replacement parts
M04-0045-B-SAV or P10-1317
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...