3.3.3.1 CALIBRATION
Beta gauge
The screen shown below displays:
the Beta calibration coefficient (K) determined during the previous calibration,
the value of the reference gauge used for the calibration,
the number of consecutive calibrations necessary to obtain a calibration value,
the date of the previous calibration.
In this screen, it is possible to manually modify the K calibration coefficient (for example, after a
calibration or a manual check): using the [
] and [
] keys, select the value of the K coefficient, then
press down the [
] key and modify the coefficient value with the [
] [
] [
] [
] keys. Then validate
with [
]
By pressing down the [»] key, the user has access to the display which allows calibration of the Beta
gauge.
It is necessary to press down the [Auto] key in the screen shown above to access the calibration.