The sampling temperature control is managed by the software of the device in the following way:
Sampling is carried out at atmospheric temperature.
A continuous measurement of external relative humidity is carried out,
When the RH value exceeds the empirical threshold of 60%, the line is heated and controlled (with
a PID standard integration) at a temperature equal to the atmospheric tempe 5 °C, in
order to avoid condensation inside the line.
T
head
= T
atmospheric
+ 5 °C.