Environnement
S.A
MP101M
Duplication prohibited
SEPTEMBER 2013
3–28
This screen displays date and hour of the last calibration, the pressure and temperature values to
which flow-rate was calibrated, and flow-rate parameters with the calculated values of the new F
coefficient.
Pressures
fields :
P1
field : pressure value upstream the critical orifice,
P2
field: pressure value downstream the critical orifice,
Atm.
field : atmospheric pressure measurement.
The atmospheric pressure measurement is performed continuously.
Temperatures
fields :
Filter
field : temperature at level filter,
Atm.
field : External temperature measured by the meteo sensor (if the meteo sensor option is
available).
Flow-rate
fields:
Reference
field : the setting value of flow is set to 1 m
3
/h, it does not vary.
Previous F coef.
field : the value of the previous factor is used to calculate the new factor.
Span
field : reference flow
New F coef.
field : obtained by calculation :
)
(
)
(
)
(
measured
previous
new
Flow
Span
K
K
To carry out calibration, press down the [>>] key to return to the previous screen, and the [Auto] key to
launch the automatic calibration of flow rate. The
« Calibration cycle »
screen shown below is
displayed.
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...