EM-2115 APPENDIX D BROADBAND CORRECTION FACTORS
(EM2110D-1)
APPENDIX D
BROADBAND CORRECTION FACTORS
FOR
ELECTRO-METRICS MODEL EM-2115 (EMC-30)
SERIAL NO: TYPICAL
RF BAND NO:
FREQUENCY
RANGE
10 kHz1
BANDWIDTH
100 kHz1
BANDWIDTH
1 MHz1
BANDWIDTH
1
9-35 kHz
+41
xxxx
xxxx
2
35-75 kHz
+39
xxxx
xxxx
3
70-150 kHz
+40
xxxx
xxxx
4
120-250 kHz
+39
xxxx
xxxx
5
250-500 kHz
+40
xxxx
xxxx
6
0.5-1.1 MHz
+40
xxxx
xxxx
7
1.1-2.4 MHz
+40
xxxx
xxxx
8
2.4-5.5 MHz
xxxx
+21
xxxx
9
5.5-12.5 MHz
xxxx
+21
xxxx
10
12.5-30 MHz
xxxx
+20
xxxx
11
25-50 MHz
xxxx
xxxx
+1
12
50-100 MHz
xxxx
xxxx
+1
13
100-200 MHz
xxxx
xxxx
+4
14
200-500 MHz
xxxx
xxxx
+4
15
500-1000 MHz
xxxx
xxxx
+4
NOTE: 1.
Widest available bandwidth for the frequency range noted.
Add the correction factor given for the frequency range in use to the front panel digital
meter indication, in dB(µV), to obtain the broadband signal level in dB(µV/MHz). The broad-
band signal amplitude level accuracy will be ±2.0 dB using the above chart factors.
QUALITY ASSURANCE:
.
Summary of Contents for EM-2115
Page 1: ...INSTRUCTION MANUAL INTERFERENCE ANALYZER MODEL EM 2115 EMC 30...
Page 68: ...EM 2115 3 0 PERFORMANCE CHECK EM2115S3 13...
Page 104: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 36...
Page 106: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 38...
Page 108: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 40...
Page 110: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 42...
Page 112: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 44...