EM-2115 3.0 PERFORMANCE CHECK
(EM2115S3-10)
200 Hz
0.100 MHz
(FR 3)
170-220 Hz
1 kHz
0.100 MHz
(FR-3)
0.9-1.1 kHz
10 kHz
0.120 MHz
(FR-4)
8-10 kHz
10 kHz
0.240 MHz
(FR-4)
8-10 kHz
10 kHz
0.260 MHz
(FR-5)
8-10 kHz
10 kHz
0.400 MHz
(FR-5)
8-10 kHz
10 kHz
1.00 MHz
(FR-6)
8-10 kHz
10 kHz
2.30 MHz
(FR-7)
8-10 kHz
10 kHz
12.0 MHz
(FR-9)
8-10 kHz
100 kHz
150.0 MHz
(FR-13)
100-110 kHz
NOTE:
FR = Frequency Range.
3.4.8 Image Rejection
Specification: 45 dB minimum
Equipment required: Signal Generator
a.
Set the front panel controls as follows:
DETECTOR ...................................................AVERAGE
SAMPLE TIME Control ................................Mid-range.
FREQUENCY RANGE .................................0.009-0.035 MHz.
FREQUENCY INDICATION .......................0.220 MHz.
BANDWIDTH ...............................................200 Hz.
ATTENUATION ...........................................40 dB
b.
Apply a 22 kHz CW signal at an input level of +60 dB(µV) (1 mV) to the EM-
2115 RF Input connector.
c.
Peak the Signal Generator for a maximum indication on the front panel analog
meter, the digital meter should in60 dB(µV) ±2.0 dB. If not, adjust the in-
put signal level. This is the reference point for the following procedure.
d.
Tune the Signal Generator to twice (2X) the IF plus the EM-2115 indicated fre-
quency.
Summary of Contents for EM-2115
Page 1: ...INSTRUCTION MANUAL INTERFERENCE ANALYZER MODEL EM 2115 EMC 30...
Page 68: ...EM 2115 3 0 PERFORMANCE CHECK EM2115S3 13...
Page 104: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 36...
Page 106: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 38...
Page 108: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 40...
Page 110: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 42...
Page 112: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 44...