EM-2115 APPENDIX E ACCESSORY DATA
(EM2115E-7)
217563, the EM-6960 is designed specifically for EMI measurements and specification compli-
ance testing.
The antenna is precision machined from aluminum. The 50
Type N connector is
mounted on the base block of the antenna. A bracket is provided that enables the antenna to be
mounted to a standard tripod.
Each Model EM-6960 Antenna is individually calibrated during the manufacturing pro-
cess. VSWR is checked for conformance to a standard characteristics chart. Apparent gain at 1.0
meter from the end of the antenna is determined and used to check conformance to the antenna
factor. This factor is used during specification compliance testing to convert receiver readings, in
dB(µV), to field intensity units (dB(µV)/m). The conversion is accomplished by adding the an-
tenna factor in dB/m to the receiver reading in dB above 1 microvolt.
Typical antenna radiation patterns (horizontal and vertical polarization) at 200 MHz, 500
MHz, 1.0 GHz, and 2.0 GHz are also included in each antenna manual.
E-2.5.3 EM-6960 Applications
The EM-6960 is designed specifically for EMI measurements and specification compli-
ance testing to MIL-STD-416 plus other government and federal standards.
E-2.6 EM-6892 Remote Active Antenna
E-2.6.1 EM-6892 Specifications
a. Electrical:
Frequency Range (Calibrated):
10 kHz to 100 MHz.
Calibration Factor Chart furnished with each antenna. Refer to Figure E.5 for
typical antenna factor chart.
3 dB Bandwidth:
9 kHz to 20 MHz.
Input Impedance:
5 M
resistive.
Sensitivity:
From -13 dB(µV/m/Hz) at 10 kHz to -40 dB
(µV/m/Hz) at 30 MHz, normalized to 1 Hz
bandwidth.
Dynamic Range:
Typically 120 dB, 1 µV/m to 1 V/m.
Saturation Level:
Typically
126
dB(µV/m)
CW,
105
dB(µV/m/MHz) Broadband.
Summary of Contents for EM-2115
Page 1: ...INSTRUCTION MANUAL INTERFERENCE ANALYZER MODEL EM 2115 EMC 30...
Page 68: ...EM 2115 3 0 PERFORMANCE CHECK EM2115S3 13...
Page 104: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 36...
Page 106: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 38...
Page 108: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 40...
Page 110: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 42...
Page 112: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 44...