Page 94 of 126
198-0853-02 Rev F
A.6.5.1
Local Calibration Menu
Enter the local calibration menu from the calibrate configuration menu. This
menu is used to calibrate the current and voltage meters and the power supply
output.
Local Calibration
Menu Page 2
Local Calibration
Menu (Pg 1 of 2)
.
Select YES to reset all
local calibration
settings to default
values.
The upper current output is calibrated at 95% of the
current rating. In operate mode the output should be
adjusted to 95% of the current rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.
.
In operate mode the upper current output calibration
value is used to output 95% of the current rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.
.
The lower current output is calibrated at 10% of the
current rating. In operate mode the output should be
adjusted to 10% of the current rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.
.
In operate mode the lower current output calibration
value is used to output 10% of the current rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.