B.1.10 CRC
Disable/Enable Command “c”
@AA.a Sync character (@) and unit address, AA = Unit ID (0 to 99; 0 is a global ID), a
= Channel ID (always 0)
c1
Command character and Type
– 0=read, 1=set, 2=activate(n/a), 3=ack, 4=nak
#1,
Number of fields character and number of fields after the comma, excluding
CRC
0,
CRC flag
– 0=disable, 1=enable. Default on power supply power-up is
Enabled.
crcf
CRC
CRC value in ascii decimal. NOTE: examples show actual CRC values.
crlf
Carriage return and Line feed (0x0D 0x0A)
Example of CRC Enable/Disable Command:
Disable CRC command
@00.0c1#1,0,32516
Enable CRC command
@00.0c1#1,1,61189
Read CRC state command.
@12.0c0#0,25165
ACK response for Read CRC State command from unit with address = 12, with CRC enabled. Shown
with and without the delimiter text enabled.
@12.0c3#1,1crcf,58566
@12.0c3#1,1,15265
Notes:
1. Default CRC setting on power supply power-up is
Enabled
.
2.
The channel ID ‘a’ should always be zero.