Page 34 of 126
198-0853-02 Rev F
powered off then when the unit is powered on it will resume the timed cycle where it left off after a
short count down period. The automatic resume can be terminated during the count down period
by pressing the operate button. If a timed cycle was paused when the unit was powered off then
the unit will power on in the paused state. If a manual cycle was running (in operate) when the
unit was powered off then the then when the unit is powered on it will go back into operate mode
after a short count down period. The return to operate mode can be terminated during the count
down period by pressing the operate button. If the power supply was in standby mode when it
was powered off then it will power on in standby mode.
See
A.2.Power On/Off Sequences
for more details on what is displayed during power on and
power off in various power on configurations.
The power on configuration can be changed in the “Unit Start System Menu”. Use the following
procedure to change the power on configuration. The “System Configuration Menu” is located
within the “Unit Configuration Menu (
see 6.3.Unit Configuration
). See
A.6.4.System Configuration
Menu
and
A.6.4.1.Unit Start System Menu
for more information regarding menu navigation.
1)
From the “Unit Configuration Menu” rotate the current encoder until the display shows
“CFg” “SyS”, and then press the voltage encoder switch to enter the “System
Configuration Menu”.
2)
From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “Strt”, and then press the voltage encoder switch to enter the “Unit Start
System Menu”.
3)
From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “CtrL”, and then press the voltage encoder switch to view and change the power
on configuration ( or
“start control” configuration).
4)
The power on configuration (or “start control” configuration) is indicated by displaying
“Unit Strt CtrL” on the top display and the selected power on configuration in the bottom
display. The SET lamp is blinking on and off to indicate that a setting is displayed.
Rotate the current encoder to change the selected power on configuration. Select
“Stby” to configure the unit to always power on in standby mode. Select “LASt” to
configure the unit to power on in the same mode it was in when it was powered off.
Once the desired configuration is selected press the voltage encoder switch to accept
the value and return to the “Unit Start System Menu”.
5)
From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “End”. Then press the voltage encoder switch to exit the “Unit Start System
Menu” and return to the “System Configuration Menu”.
6)
From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “End”. Then press the voltage encoder switch to exit the “System
Configuration
Menu” and return to the “Unit Configuration Menu”.
7) If any settings were changed
in the “System Configuration Menu” you will be prompted
to store the changes. Select “yES” to accept the changes or “no” to cancel the
changes.
6.3.5. Power Fail Alarm Configuration
The power supply can be configured to report error codes if power is interrupted while the supply
is in operate mode. If the power failure alarms are enabled the appropriate error code (error
codes 102 or 103) is set when power fails during operate mode. The error code will remain active