The temperature sensor value is
outside the valid range of
temperatures. Failed temperature
sense circuit or unit is located in an
environment where the ambient
temperature is outside the
recommended temperature range.
Verify that the rectifier is located
where ambient temperatures do
not exceed the recommended
temperature range. Power off
the rectifier and allow it to return
to ambient temperature. If fault
re-occurs report the condition to
Dynatronix.
YES
YES
76
X
Modules detected at power up do not
match the module configuration
stored in EEPROM
There is either a problem with
the configuration, or the incorrect
power modules have been
installed.
YES
NO
77
X
One or more calibration values could
not be stored to EEPROM.
Re-calibrate. If the fault persists
contact Dynatronix.
NO
YES
78
X
N/A
N/A
N/A
N/A
79
X
X
N/A
N/A
N/A
N/A
80
X
X
N/A
N/A
N/A
N/A
81
X
N/A
N/A
N/A
N/A
82
X
N/A
N/A
N/A
N/A
83
X
N/A
N/A
N/A
N/A
84
X
N/A
N/A
N/A
N/A
85
X
N/A
N/A
N/A
N/A
86
X
N/A
N/A
N/A
N/A
87
X
N/A
N/A
N/A
N/A
88
X
N/A
N/A
N/A
N/A
89
X
N/A
N/A
N/A
N/A
90
X
N/A
N/A
N/A
N/A
91
X
N/A
N/A
N/A
N/A
92
X
N/A
N/A
N/A
N/A
93
X
N/A
N/A
N/A
N/A
94
X
X
Current over tolerance alarm (CH1).
The difference between the current
reading and the current setting on
CH1 exceeded the tolerance setting
(reading greater than the current
setting).
Out of tolerance alarms self clear
at the start of a new cycle.
Optionally out of tolerance
alarms can be manually cleared
at any time, but will be set again
if the condition persists.
NO
YES
95
X
X
Current over tolerance alarm (CH2).
The difference between the current
reading and the current setting on
CH2 exceeded the tolerance setting
(reading greater than the current
setting).
Out of tolerance alarms self clear
at the start of a new cycle.
Optionally out of tolerance
alarms can be manually cleared
at any time, but will be set again
if the condition persists.
NO
YES
96
X
X
Current under tolerance alarm (CH1).
The difference between the current
reading and the current setting on
CH1 exceeded the tolerance setting
(reading less than the current
setting).
Out of tolerance alarms self clear
at the start of a new cycle.
Optionally out of tolerance
alarms can be manually cleared
at any time, but will be set again
if the condition persists.
NO
YES
97
X
X
Current under tolerance alarm (CH2).
The difference between the current
reading and the current setting on
CH2 exceeded the tolerance setting
(reading less than the current
setting).
Out of tolerance alarms self clear
at the start of a new cycle.
Optionally out of tolerance
alarms can be manually cleared
at any time, but will be set again
if the condition persists.
NO
YES