10.1.3. Restoring the default calibration values
The default calibration values can be restored with the following procedure.
CAUTION:
Restoring the default calibration values will cause any previous calibration values to be
overwritten. A calibration must be performed after loading the default calibration points by
following the calibration procedure in order to get accurate output levels.
1.
From the “Unit Configuration Menu” rotate the current encoder until the display shows
“CFg” “CAL”. Press the voltage encoder switch to enter the “Calibrate Configuration
Menu”. See
A.6.Unit Configuration Menu
and
A.6.5.Calibrate Configuration Menu
for more
details on menu navigation.
2.
From the “Calibrate Configuration Menu” rotate the current encoder to select the desired
calibration menu. Select “LCL” for the local current and voltage meters and current and
voltage output. Select “OPt” for the analog option card inputs and outputs.
3.
From the calibration menu rotate the current encoder until the display shows “dFLt” on the
bottom display. Press the voltage encoder switch to proceed with loading default
calibration values.
4. The default calibration confirmation screen show
s “dFLt” on the top display and “yES” or
“np” on the bottom display. Rotate the current encoder to select “yES” and press the
voltage encoder switch to load the calibration default values.
5. Proceed with a normal calibration.