Smart LT Manual
Dynatronix, Inc.
198-0853-02 Rev F
Page 39 of 126
The password can be changed from the “Unit Configuration Access Code System Menu”. Use the
following procedure to change the unit configuration password (also referred to as the unit
configuration access code). The “System Configuration Menu” is located within the “Unit
Configuration Menu (
see 6.3.Unit Configuration
). See
A.6.4.System Configuration Menu
and
A.6.4.5.Unit Configuration Access Code System Menu
for more information regarding menu
navigation.
1)
From the “Unit Configuration Menu” rotate the current encoder until the display shows
“CFg” “SyS”, and then press the voltage encoder switch to enter the “System
Configuration Menu”.
2)
From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “COdE”, and then press the voltage encoder switch to enter the “Unit
Configuration Access Code System Menu”.
3) The first scr
een in the “Unit Configuration Access Code Menu” shows the existing
access code value. To change the access code rotate the current encoder to change
the selected digit (blinking digit). Each press of the current encoder switch advances
the selected digit. After the desired code is entered press the voltage encoder to accept
the new access code.
4) If the access code is changed an informational warning message is displayed to
indicate that the code has been changed. This screen displays “CHng” on the top
di
splay and “COdE” on the bottom display. This screen is displayed for a short period of
time and can be skipped by pressing the voltage encoder switch.
5) After the warning message is displayed a confirmation screen is displayed. The top
display shows “CHng COdE” and the bottom display shows the confirmation selection.
Rotate the current encoder to change the confirmation selection. Select “yES” to accept
the access code change or “no” to cancel the change. Press the voltage encoder
switch to accept the se
lection and exit the “Unit Configuration Access Code System
Menu” and return to the “System Configuration Menu”.
6)
From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “End”. Then press the voltage encoder switch to exit the “System
Configuration
Menu” and return to the “Unit Configuration Menu”.
7) If any settings were changed
in the “System Configuration Menu” you will be prompted
to store the changes. Select “yES” to accept the changes or “no” to cancel the
changes.
You must select “yES” to accept the changes to the access code.
6.3.10. Setting Unit Defaults
All
settings in the “Unit Configuration Menu” and the “Unit Setup Menu” can be reset to default
values from the “Set Defaults System Menu”. Calibration values are not affected.
Use the following procedure to reset all settings to default values. The “System Configuration
Menu” is located within the “Unit Configuration Menu (
see 6.3.Unit Configuration
). See
A.6.4.System Configuration Menu
and
A.6.4.6.Set Defaults System Menu
for more information
regarding menu navigation.