4/Tests and Troubleshooting
1503-0151-000
5/26/0
4-31
Calibrate Bleed Resistor
The software calibrates the bleed resistor using the following procedure steps (be-
sides the setup procedure this calibration is completely automated). First the soft-
ware finds the value at which to set the flow valve to get a manifold pressure of 91-
92 cmH
2
O. Once it finds this starting point value the automated calibration begins.
Values shown here are for example only and are not necessarily typical.